MTL K K6050 & KG6050 Portable Gas Analysers

  • Wide range of gases monitored, including hydrogen, carbon dioxide, argon and SF6
  • Optional oxygen measurement
  • Rugged and portable
  • Fast warm-up with rapid response time
  • Weatherproof to IP67 Applications
  • Hydrogenation processes
  • Gas purity
  • Food processing
  • Refrigeration systems
  • Power generation/distribution
  • Breweries
  • Metallurgical atmospheres
  • Products of fermentation
Request for quote
Description

Description

The K6050 range gas analyzers accurately measures the changes of one constituent in binary or pseudo-binary gas mixture, i.e. mixtures in which only one constituent changes. Examples include hydrogen, carbon dioxide, argon, helium  and various halogenated hydrocarbons such as the Freons. Almost any single constituent of a gas mixture can be measured providing its thermal conductivity  differs from that of the other components. Ranges from high ppm to 100% are possible depending on the gas being measured.

Specifications
  • Display

    Dot-matrix LCD showing two or four lines
    of alphanumeric characters

  • Thermal conductivity ranges and gas types

    0 to 20%, to 0 to 100%, 80 – 100% (alter
    ranges on application)
    Carbon dioxide, argon,methane, SF6
    0 to 1%, to 0 to 5%, 0 to 20%, 0 to 100%
    Hydrogen, helium
    Consult MTL product line for other gases
    and ranges

  • Accuracy

    ±1% of span (typically)

  • Stability

    Better than 1% fsd/month

  • Oxygen measurement ranges (KG6050 only)

    0 to 100%
    Resolution: 0.1%
    Accuracy: ±0.25%
    0 to 25%
    Resolution: 0.1%
    Accuracy: ±0.1% (5 to 100% of scale)
    ppm level ranges also available, subject
    to application

  • Oxygen cell life

    Typically 2 years

  • Speed of response (typical)

    Oxygen measurement: 12s (T90)
    Thermal conductivity measurement: 20s

  • Sample connections

    Compression fitting suitable for 0.25 inch (6mm) OD
    tubing or quick release connector

  • Sample pressure

    Pump off: +6 bar G maximum
    Pump on: –100mb G minimum
    Higher pressures available on request

  • Sample flow

    100 to 300ml/min for optimum performance

  • Sample temperature

    –10 to +55°C (non-condensing)

  • Outputs

    0 to 1V analogue

  • Ambient temperature

    –5 to +40°C

  • Battery capacity

    6 hours

  • Battery chargers – supplied

    Mains: 240V ac (UK only)
    Vehicle cigarette lighter: 12V dc

  • Case

    Tough co-polymer resin with carrying handle

  • Dimensions

    338mm wide x 295mm deep x 162mm high with
    handle parked in forward position

  • Weight

    5.2kg approximately