MTL K K6050 & KG6050 Portable Gas Analysers
  

·         Wide range of gases monitored, including hydrogen, carbon dioxide, argon and SF6

·         Optional oxygen measurement

·         Rugged and portable

·         Fast warm-up with rapid response time

·         Weatherproof to IP67 Applications

·         Hydrogenation processes

·         Gas purity

·         Food processing

·         Refrigeration systems

·         Power generation/distribution

·         Breweries

·         Metallurgical atmospheres

·         Products of fermentation

Datasheet:  


Request for quote

Description

The K6050 range gas analyzers accurately measures the changes of one constituent in binary or pseudo-binary gas mixture, i.e. mixtures in which only one constituent changes. Examples include hydrogen, carbon dioxide, argon, helium  and various halogenated hydrocarbons such as the Freons. Almost any single constituent of a gas mixture can be measured providing its thermal conductivity  differs from that of the other components. Ranges from high ppm to 100% are possible depending on the gas being measured.

Specifications

 

Display

Dot-matrix LCD showing two or four lines
of alphanumeric characters

Thermal conductivity ranges and gas types

0 to 20%, to 0 to 100%, 80 – 100% (alter
ranges on application)
Carbon dioxide, argon,methane, SF6
0 to 1%, to 0 to 5%, 0 to 20%, 0 to 100%
Hydrogen, helium
Consult MTL product line for other gases
and ranges

Accuracy

±1% of span (typically)

Stability

Better than 1% fsd/month

Oxygen measurement ranges (KG6050 only)

0 to 100%
Resolution: 0.1%
Accuracy: ±0.25%
0 to 25%
Resolution: 0.1%
Accuracy: ±0.1% (5 to 100% of scale)
ppm level ranges also available, subject
to application

Oxygen cell life

Typically 2 years

Speed of response (typical)

Oxygen measurement: 12s (T90)
Thermal conductivity measurement: 20s

Sample connections

Compression fitting suitable for 0.25 inch (6mm) OD
tubing or quick release connector

Sample pressure

Pump off: +6 bar G maximum
Pump on: –100mb G minimum
Higher pressures available on request

Sample flow

100 to 300ml/min for optimum performance

Sample temperature

–10 to +55°C (non-condensing)

Outputs

0 to 1V analogue

Ambient temperature

–5 to +40°C

Battery capacity

6 hours

Battery chargers – supplied

Mains: 240V ac (UK only)
Vehicle cigarette lighter: 12V dc

Case

Tough co-polymer resin with carrying handle

Dimensions

338mm wide x 295mm deep x 162mm high with
handle parked in forward position

Weight

5.2kg approximately

Datasheet MTL K K6050 & KG6050 Portable Gas Analysers

Download the datasheet below.